12.5.3.3 Replacing the AC Source Motion Board

Procedure
Before starting to replace the AC source
motion board:
1. Dismount the transmission scanner from
the detector (see section 12.5.1.1, page 12-
29).
2. Detach the line source compartment from
the bracket (see section 12.5.1.2, page 12-
33).
3. Remove line source compartment covers
(see page 12-41).
4. Detach the connecting cable (see section
12.5.3.2, page 12-43).
When all of the above steps have been
performed, remove the collision detection
shutter end cover:
5. Use a T08 Torx key to remove the 3
screws indicated in the photo on the right.
6. Remove the cover.
12-44 CardioMD Service Manual
Details
1MAN0158-H11 04-2005 CSIP Level 1