Gantry
The function of the switches shown in Figure 3.8 is as follows:
Switch
Collision
Override
Ultimate
Stop
Override
Ultimate
Stop
reset
Zero
Enable
Zero R
Zero H
Zero V
Table 3.3 The function of the safety board switches
3.2.3

Motion Controllers

3.2.3.1

Motions

The CardioMD system has four independent motions:
DETECTOR ROTATE
DETECTOR Y
DETECTOR X
TABLE LONGITUDINAL
3-10 CardioMD Service Manual
Designation
Type
COLLISION
Slide
OVERRIDE
switch
ULT_STOP
Slide
OVERRIDE
switch
ULT_STOP
Push
RESET
button
ZERO_ENA
Push
button
ZERO_R
Push
button
ZERO_H
Push
button
ZERO_V
Push
button
The gear carrying the detectors can rotate over a range of 270°
The detectors can move vertically up and down over a range of
962 mm
The detectors can move horizontally towards and away from the
patient table over a range of 495 mm
The total travel range for the patient table is 100 mm.
Description
This switch function is for manufacturing and service purposes.
When the switch is ON (position downwards), the collision
sensing on the collimators and detectors is disabled. At the same
time, the adjacent red LED is lit, indicating that a safety function
is disabled.
Caution. Make certain that the switch is OFF (position upwards)
when the system is in normal use
This switch disables the bi-stable Ultimate Limit relay output. If
a motion has passed its ultimate end stop, the Limit relay
interrupts power to the system and the Ultimate Stop Override
has to be applied in order to re-enable motor power to be able to
back out. Use the Collision Override button and the appropriate
motion button on the hand controller.
An adjacent red LED is lit when the switch is ON.
Caution. Make certain that the switch is OFF (position upwards)
when the system is in normal use
This button resets the bi-stable Ultimate Limit relay output. The
relay must be reset after clearance of the motion at the ultimate
stop limit
This button is used to set the calibration points for the absolute
encoders for the rotate, horizontal and vertical motion calibration
points.
The button must be pressed simultaneously with each of the
motion specific set buttons (Zero R, Zero H, Zero V)
This button sets the absolute encoder calibration point for the
rotate axis when pressed simultaneously with the Zero Enable
button
This button sets the absolute encoder calibration point for the
horizontal (X) axis when pressed simultaneously with the Zero
Enable button
This button sets the absolute encoder calibration point for the
vertical (Y) axis when pressed simultaneously with the Zero
Enable button
1MAN0158-H11 04-2005 CSIP Level 1