Procedure
8. Connect the cable from the transmission
scanner to the socket on the detector as
shown in the photo on the right:
Insert the cable plug into the socket.
Tighten the screws to secure the
connector.
While the cable is connected, the handle
cannot be moved to the unlocked position.
9. Repeat steps 5 to 8 to mount and connect
transmission scanner 2 on detector 2.

12.6.4 Software Installation

12.6.4.1 Upgrading EDC Board Firmware

The CardioMD AC option requires EDC board firmware 7CTL0095-E05 or later. The system's
current EDC board firmware revision is recorded in the system log file. Check the most recent log
file in the directory
Tools Required
CardioMD™ installation Disk. P/N: 9CSY0906.
Procedure
Refer to Chapter 7 Software Update Procedures, the section Upgrading EDC Firmware for
instructions.

12.6.4.2 Installing AC Option Acquisition Software

The CardioMD software is shipped on a CD-ROM that includes two batch files used to update or
perform a complete reinstallation of the acquisition software. To install the AC Option software,
perform an update of the acquisition software as described below.
Tools Required
CardioMD™ installation Disk. P/N: 9CSY0906.
Note. As a precaution, before updating the CardioMD software, perform a back-up of all user data
on the system, i.e. detector calibrations, user-defined protocols etc. See the Appendix of the
CardioMD Operator's Manual for instructions.
Note. As a precaution, ensure that the previous software version CD-ROM is available in order to
be able to restore the original version of the acquisition software.
1MAN0158-H11 04-2005 CSIP Level 1
Details
C:\cardiocam\SystemLog
.
CardioMD Service Manual 12-105
CardioMD AC Option