Repair Procedures
Procedure
The Static page appears.
11. Click on the Pencil button to access
energy setup.
12. For Eset1, enter data for energy window 1
(Ew1):
Peak keV: 100.0
Width %: 20.0
13. Leave Eset2 and Eset3 blank.
14. Click Done.
The Isotope setting is not important.
15. Click Start.
Now you need to open the line source shutter
of the transmission scanner that you are
aligning. Proceed as follows:
12-84 CardioMD Service Manual