Installation
Connect the PC-gantry cable to the gantry as follows:
Procedure
1. Push the cable coming from the
acquisition PC wall-mount through the
entry hole in the gantry.
2. Fasten the plate to the gantry using the
M4 x 8 screws.
3. Connect the hand-controller/LED/E-stop
cable to the DB25 connector on the
internal connector panel.
4. Connect the power cable to the IEC
connector.
5. Connect the FireWire cable to any of the
ports on the FireWire hub.
6. Attach the eye terminal of the
yellow/green ground wire to the ground
screw on the power supply frame.
8-58 CardioMD Service Manual
Details
DB25 connector
IEC connector
FireWire hub
Gound wire
1MAN0158-H11 04-2005 CSIP Level 1