Installation

8.3.3.18 Leveling the Table Base

Procedure
1. Remove the mounting screws in the rear
end of the table base.
2. Using a level and the adjusting tool, level
the table base.
3. Remount the mounting screws.

8.3.3.19 Removing the 2 Shipping Locks

Procedure
1. Tighten the mounting screws connecting
the table base and gantry frame.
2. Using an 6 mm Allen wrench, remove the
2 shipping locks located in the gantry
sidebars.

8.3.3.20 Mounting the Patient Table Support

Procedure
1. Mount the table end support onto the
gantry base using two M8 bolts.
8-38 CardioMD Service Manual