Repair Procedures
Procedure
19. Mount the two Allen screws holding the
side rail to the compartment.
20. Mount the two Allen screws holding the
source motion board.
Note. It is important to fasten the side rail
before mounting the Allen screws in the
source motion board.
21. Mount the actuator bracket, inserting the
anchor into the solenoid.
Note. Make sure that the end of the bracket
fits into the groove in the side of the line
source compartment.
22. Re-insert the shutter indicator / manual
lever.
23. Mount the four Allen screws holding the
indicator and the actuator bracket to the
slide bearing.
Note. Do not tighten the Allen screws yet.