Procedure
3. Connect the SW load cable to the motion
controller serial port labeled BZ2 RS232.
This does not require that any existing cables
be removed.
If you are using a laptop acquisition PC:
4. Use the USB to RS 232 adapter to connect
the opposite end of the SW load cable to
one of the acquisition PC's USB ports.
See section 7.2.2, page 7-2 for instructions on
determining which COM port you are using.
If you are using a tower acquisition PC:
5. Connect the opposite end of the cable to
the acquisition PC's COM1 serial port.
Note. Downloading software to the motion
controller does not require the use of a NULL
modem cable.
6. Disconnect the CANbus cable between the
two motion controllers by unplugging the
cable connected to CAN_IN on the slave
motion controller (controller to the left).
7. Insert the software installation CD in the
CD-ROM drive.
To download software to the motion control-
ler, you must start the Flash program. Proceed
as follows:
8. In the Windows Explorer, select the folder
on the CD-ROM.
MC
9. In this folder, double-click the icon
to start the Flash program.
flash.exe
1MAN0158-H11 04-2005 CSIP Level 1
Details
Software
load cable
CANbus
cable
III
I, II
Software Update Procedures
CardioMD Service Manual 7-15