Installation Procedures

9.3 Installation Procedures

9.3.1

Collimators

Collimators for the CardioMD system are mechanically coded for detector 1 or detector 2. When
receiving a spare collimator, you may have to change the mechanical coding of the collimator to
fit it onto the detector.The procedure describes the steps involved when replacing a collimator.
Procedure
1. Determine if the collimator is to be coded
for detector 1 or detector 2.
2. If the collimator is to be coded for detector
1, place the two M4x6 screws to the right
of the collimator locks, as shown in the
photo on the right.
3. To code the collimator for detector 2,
place the two M4x6 screws to the left of
the collimator locks, as shown in the photo
on the right.
9-36 CardioMD Service Manual
Details
1MAN0158-H11 04-2005 CSIP Level 1