Safety
1.3.7

Emergency Stop

Pressing the Emergency Stop button located on the side of the gantry or the one located next to the
acquisition PC will stop and disable all motion.
Emergency Stop
Figure 1.1 Emergency Stop buttons
Note. The appearance of your system may be slightly different from the systems shown in the photographs of this
manual. However, there is no difference in functionality
Note. E-stop buttons latch mechanically when pressed. To release an E-stop button, turn the
button in the clockwise direction.
Pressing an Emergency Stop button causes the current acquisition to stop. Any acquisition motion,
for example tomography, will cease.
When the Emergency Stop button is released (and power to the rest of the system has not been
interrupted), you can enable hand controlled motion by pressing the hand controller Collision
Override Button once. An interrupted acquisition remains stopped until you instruct the system to
resume or restart the acquisition.
CardioMD Service Manual 1-5
1MAN0158-H11 04-2005 CSIP Level 1