NEMA Multiple Window Spatial Registration (MWSR)
Procedure
After the last 400 kilocounts the result pops
up.
16. Note the result.
The MWSR specification is ≤ 2.0 mm.
17. Remove the GEOCAL mask and the re-
maining lead pig from the detector and
mount the collimator.
18. Click Exit to return to the NEMA menu.
Note. To measure the other detector, place the CardioMD system in the patient load position be-
fore repeating the procedure outlined above.
10-14 CardioMD Service Manual
Details
1MAN0158-H11 04-2005 CSIP Level 1