Philips CARDIOMD Service Manual page 111

Table of Contents

Advertisement

COLLIS. OK
COLLISION
COLLIS.
DISABLED
ULT_STOP
DISABLED
ROT.
ULT_OK
HORIZ.
ULT_OK
VERT.
ULT_OK
E_STOP OK
TABLE
E_STOP ON
REMOTE
E_STOP ON
RUNAWAY
1MAN0158-H11 04-2005 CSIP Level 1
This green LED will be on during normal operation. During collision the LED is
turned off.
This red LED is on during collision and off during normal operation.
This red LED is on when the Collision Override switch is on (position down) to
warn that the collision circuitry is disabled.
This red LED is on when the Ultimate Stop Override switch is on (position down)
to warn that the ultimate stops are disabled.
This green LED is on during normal operation. The LED is turned off when the
ultimate stop switch on the rotational motion has been triggered (motion limit
exceeded), or when either the HORIZ. ULT_OK or VERT. ULT_OK has been
triggered.
This green LED is on during normal operation. The LED is turned off when the
ultimate stop switch on the horizontal (X) motion has been triggered (motion limit
exceeded), or when VERT. ULT_OK has been triggered.
This green LED is on during normal operation. The LED is turned off when the
ultimate stop switch on the vertical (Y) motion has been triggered (motion limit
exceeded).
This green LED is on during normal operation. The LED is turned off when one of
the Emergency Stop buttons has been pressed.
This red LED is off during normal operation. The LED is turned on when the
gantry E-Stop button has been activated.
This red LED is off during normal operation. The LED is turned on when the E-
Stop button next to the acquisition PC has been activated.
This red LED is off during normal operation. The LED is turned on when one of
the motion controllers has issued a Runaway signal.
Diagnostics
CardioMD Service Manual 5-7