Gantry Software Updates
Procedure
not, continue with the instructions below.
After upgrading the EDC board's micro con-
troller firmware, the ECG connection must be
restored and the EDC module must be power
cycled:
14. Remove the serial interface cable from the
connector A1 RS232 and reconnect the
ECG cable (see step 2 in the section
Preparing for Firmware Upgrade on page
7-8).
15. Remove the EDC power cable connected
to connector A4 POWER. Reconnect the
power cable.
Proceed to upgrade the micro controller firm-
ware of the other EDC board:
16. Connect the serial interface cable to the
other EDC board as described on page 7-8.
17. Repeat steps 1 (page 7-8) to 15 (page 7-
10) of this procedure to update the micro
controller firmware of the EDC module on
the other detector.
18. When both EDC modules have been up-
dated, remount the detector top cover fol-
lowing instructions provided in Chapter 6.
Repair Procedures.
Upgrading Logic Firmware
Procedure
To download logic firmware to the EDC
board, you use the terminal emulator program
ttermpro. Proceed as follows:
1. In the Windows Explorer, select the folder
EDC CardioMD programming for
Module rev. 9CSY0900-B02 or
later\TTERMPRO
2. In this folder, double-click the icon
to start the ttermpro pro-
ttermpro.exe
gram.
7-10 CardioMD Service Manual
on the CD-ROM.
Details
Connector A4 POWER
on the EDC board
Details
1MAN0158-H11 04-2005 CSIP Level 1