Procedure
6. Use a part of the rubber protector supplied
with the system to protect the cable from
the wall to the gantry.
3 m (10 ft) of rubber cable protectors are
provided to cover cables running from the
gantry to the wall (power and network cables)
and the cable between gantry and PC. The
cable protectors should be applied where
traffic across cables must be expected.
7. Cut cable protectors to desired lengths.
Open the membrane at the base of the
protector to slide in the cable.
Connect the gantry-PC interface cable to the gantry as follows:
Procedure
8. Push the cable coming from the
acquisition PC through the entry hole in
the gantry.
9. Fasten the plate to the gantry using the
M4 x 8 screws.
10. Connect the hand-controller/LED/E-stop
cable to the DB25 connector on the
internal connector panel.
11. Connect the power cable to the IEC
connector.
12. Connect the FireWire cable to any of the
ports on the FireWire hub.
1MAN0158-H11 04-2005 CSIP Level 1
Details
Details
DB25 connector
IEC connector
FireWire hub
Installation
CardioMD Service Manual 8-63