(b) Temperature calibration
If there is a difference between the indicated
temperature of the thermostat and the actual
temperature of the liquid in a sample vial, it can be
corrected by setting of temperature calibration.
In case the indicated temperature of the thermostat is
10.0 °C and the actual temperature of the liquid in a
sample vial as measured with a calibrated
thermometer is 10.5 °C, the difference correction value
(offset value) is +0.5. So, setting is required as
follows.
STEP RACK VIAL VOL(µL) INJ
1
1
1
THERMO
THERMO UNIT (SET TEMP=1, CALIBRATION=2,
2
TOLERANCE=3, SERIAL NO.=4)
2 ENT
SELECT MODE (+OFFSET=1, -OFFSET=2)
1
Select on which side there is a difference, plus side or minus
side.
e.g.) 1 ENT
SET +OFFSET (0.0-5.0) °C
0.0
Set an offset value for correction.
e.g.) 0.5 ENT
OFFSET (°C)
0.5
ENT or ESC
Monitor screen reappears.
3 - 54
TIME
10.0
1/1
0.0
The presently set value is indicated.
OFFSET(°C)= 0.0
[I]