3.2 UTLTY Setting

3.2.1 Offset setting

The UTILITY key sets the following conditions.
• OFFSET
• TIME CONSTANT
• USE PROG (whether to use time program or not)
• PMT VOLT (photomultiplier voltage)
• OTHERS
The OFFSET is a set input signal amount that is added to the
actually measured value for output.
SET UP
UTLTY
(1)
Press the
(2)
The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_1
USE PROG=3, PMT VOLT=4, OTHERS=5)
Select OFFSET.
Press numeric key 1 and then the ENT key.
(3)
The following display will be presented.
<Offset setting screen>
OFFSET (0 - 1000)
_ 0
Set an offset value.
Press numeric key (0 to 1 0 0 0, usually input 0) and then
the ENT key.
* An offset value is settable within a range from 0 to 1000 in
increments of 1 step.
* Default value: 0