Hitachi CHROMASTER 5160 Instruction Manual page 181

Table of Contents

Advertisement

Mobile-phase
liquid
Detergent
Chromaster
organizer
Chromaster
5110/5160 pump
Degassing unit
Appended Fig. 1-11 Tubing Diagram of Low-Pressure
Mixer
4-way joint
Proportioning
valve
Gradient System (stacked in two
blocks)
Connect the inlet tube from the mobile-phase container to
the IN port of degassing unit. Connect the OUT port of
degassing unit and the IN port of proportioning valve.
Connect the OUT port of the proportioning valve in the low-
pressure gradient unit to the 4-way joint. Then, connect
the purge valve of pump to port 4 of the injection valve in
the autosampler. Also, connect port 5 of the injection
valve to the inlet of column. To the outlet of column,
connect the inlet tube of detector cell. Connect the
detergent container to the wash pump of autosampler
through the degassing unit.
APPENDIX 1 - 16
Chromaster 5440
fluorescence detector
Detector cell
Chromaster 5310
column oven
Column
Injection valve
Chromaster 5210/5260
autosampler
Wash pump
Drain