SAFETY SUMMARY
WARNING Precautions in the Manual (Continued)
Electric Shock due to Contact with Inside of Instrument
When removing the pump cover for part replacement, etc., there is a risk of electric shock.
Be sure to turn OFF the power switch and disconnect the power cord from the receptacle
in advance.
Electric Shock due to Improper Grounding
This instrument is designed in conformity with the specifications of Class I in Annex H of
the IEC 61010-1 (International Electrotechnical Commission Standards) – Issue 1.
To prevent an electric shock hazard, provide a proper grounding connection.
Be sure to use the grounded 3P power cable, which is supplied as a standard
(a)
accessory for the instrument.
The use of a different type of power cable may result in an electric shock hazard.
Connect the 3P power cable to a grounded 3P power outlet.
If a grounded 3P power outlet is not available, use a grounded 3P table tap or a
(b)
3P-2P adapter. In this case, be sure to provide proper grounding connection.
For grounding connection, use a screw having a diameter of M4 or higher and a
turn count of 3 or more in threading, and a wire having a thickness of 1.25 mm.
Inflammation or Injury due to Toxic, Corrosive or Stimulative Solvent
When using a toxic, corrosive or stimulative solvent, be careful not to incur a physical
inflammation or injury. For details of the properties of each solvent and how to handle it,
refer to the relevant Material Safety Data Sheets (MSDS). Be sure to handle each
solvent properly.
Wear proper protective clothes (e.g., safety goggles) so that a solvent will not
(a)
come into direct with the skin.
(b) Ventilate the laboratory room adequately to prevent accidental inhalation of
(b)
harmful solvent vapor.
WARNING
SAFETY - 4