Pump Motor Control Setting - Hitachi CHROMASTER 5160 Instruction Manual

Table of Contents

Advertisement

3.2.2 Pump Motor Control Setting

The automatic flow rate control system (STANDARD) and the
fixed flow rate control system (FIX) are available as the pump
motor control method.
Monitor screen
SET UP
UTLTY
UTLITY
(CONDITION=1, UNIT=2,
1
EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
Select CONDITION
1
ENT
CONDITION SET
2
PURGE=3,
Select MOTOR-CTRL
2
ENT
MOTOR CONTROL
1
Select STANDARD
1
ENT
Monitor screen
The motor control method is set as STANDARD.
Select FIX
2
ENT
FIX STEP AREA
100
Monitor screen
NOTE: STANDARD is selected for ordinary use.
3 - 19
FLUSH=3
(Z-ADJ=1,
MOTOR-CTRL=2
CALIBRATION=4,
(STANDARD=1,
FIX=2)
(0,
30 - 300)
Enter FIX STEP AREA .
Input range is from 0 or 30 to 300.
ENT
LPGR=5)