3.2 UTLTY Setting
UTILITY (LEAK SENSOR=1, ANALOG OUT=2,
4
SELECT MODE (+OFFSET=1, -OFFSET=2)
1
SET +OFFSET (0.0 - 5.0) °C
0.5
OFFSET (°C)
0.5
SET UP
UTLTY
ERROR OUT=3, CALIBRATION=4)
4 ENT
OFFSET (°C)=-1.0
Select either plus or minus
side for deviation.
1 ENT
Set an offset value for the
correction.
0.5 ENT
ENT or ESC key
Monitor screen
NOTE: 1. For conducting temperature calibration, be sure to
use a calibrated thermometer for measuring
temperature.
2. There is a slight difference between the set and
actual temperatures depending on the temperature
setting. In temperature calibration, it is
recommended to determine the deviation at the set
temperature that is used most frequently.
3 - 14
Present setting is
indicated.