• System Tubing
Wash fluid
Chromaster
organizer
Chromaster
5110 pump
Degassing unit
In the Chromaster series, the inlet tube up to the pump and drain
tube are fastened to the right/left side face of each unit with tube
clamps. And the tubes are connected from the pump to the
autosampler, column and detector through the central hole of
each module unit.
Drain tray
(1)
Isocratic System
Mobile
phase
Appended Fig. 1-8 Isocratic System
Connect the inlet tube of the mobile-phase container to the
IN port of the degassing unit. Connect the OUT port of the
degassing unit to the inlet of the pump. Connect the drain
valve of the pump to port 1 of the injection valve of the
autosampler. Connect port 2 of the injection valve to the
column inlet. Connect the column outlet to the inlet tube of
the detector cell. For detergent, provide connection to the
three way solenoid valve of the syringe valve of the
autosampler through the degassing unit.
APPENDIX 1 - 12
Chromaster 5440
fluorescence detector
Detector cell
Chromaster 5310
column oven
Column
Injection valve
Chromaster 5210
autosampler
Wash pump
Drain