4.4 Troubleshooting
4.4.1 Troubleshooting of Column Oven Main Unit
Symptom
No.
(Error message)
1
ROM ERROR
PRESS "CL" KEY TO
CLEAR MESSAGE
2
RAM ERROR
PRESS "CL" KEY TO
CLEAR MESSAGE
3
PARAMETER ERROR
PARAMETERS ARE
SET TO DEFAULT
VALUE
4
LOG INFORMATION
ERROR
LOG INFORMATION IS
INITIALIZED
5
EEPROM ERROR
INSTRUMENT INFO. IS
INITIALIZED
6
DATA BUFFER
OVERFLOW
PRESS "CL" KEY TO
CLEAR MESSAGE
7
CHECK TIME
PROGRAM (TEMP)
PRESS "CL" KEY TO
CLEAR MESSAGE
8
TEMP CONTROL OFF
BY EXTERNAL ERROR
PRESS "CL" KEY TO
CLEAR MESSAGE
9
TEMP CONTROL OFF
BY LEAKAGE (GAS)
PRESS "CL" KEY TO
CLEAR MESSAGE
10
TEMP CONTROL OFF
BY LEAKAGE (LIQUID)
PRESS "CL" KEY TO
CLEAR MESSAGE
11
ABNORMAL
TEMPERATURE (OVEN)
ABNORMAL
TEMPERATURE
(PELTIER)
If a fault is noticed in the Chromaster 5310 column oven main
unit, take a remedy with reference to the following table.
Cause
Erroneous operation of
instrument or circuit board
Error in communication
Temperature specified in time
program exceeds upper
temperature limit of
-5 °C.
Temperature control has been
stopped due to external
contact input.
1. Solvent leaks from flow
path in column oven
2. Solvent remains in oven
3. Alarm level of gas leak
sensor is too low
1. Solvent leaks from flow
path in column oven
2. Solvent remains in oven
3. Contamination is attached
to liquid leak sensor.
1. Temperature sensor
abnormal
2. Circuit board faulty
4 - 11
Remedy
If error recurs after power is restored,
contact your nearest service office of
Hitachi High-Technologies
Corporation sales representative.
Check connection with chromato-
data station (CDS) and GUI
controller. And then try reconnection
with CDS.
Adjust upper temperature limit setting
or temperature setting in time
program.
Press the CL key for clearing an error
indication.
1. Wipe up leaked liquid, and
retighten tube connections.
If gas leak error still occurs,
replace tubing.
2. Eliminate all of remaining solvent.
3. Adjust alarm level of the gas leak
sensor.
1. Wipe up leaked liquid, and
retighten tube connections.
If liquid leak error still occurs,
replace tubing.
2. Eliminate all of remaining solvent
in drain tray.
3. Wipe up contamination attached
in liquid leak sensor.
If error recurs after power is restored,
contact your nearest service office of
Hitachi High-Technologies
Corporation sales representative.
4.3.2