Lamp house
Transport
screw
Appended Fig. 1-4 Transport Screws and Mounting Xenon
(4)
(5)
NOTE: After removing the transport screws, please store
Lamp
Install the xenon lamp in the lamp house. And mount it to
the instrument (see ⑧ to ⑪ in section 6.2.1 (2).)
Adjust the position of the xenon lamp.
The xenon lamp position had drastically moved to the
downward of the vertical direction to fix the transport screws.
For this reason, adjust VERTICAL first (see section 6.2.1
(3).)
them.