Periodic Maintenance; Liquid Leak Check; Washing The Flow Path; Cleaning - Hitachi CHROMASTER 5160 Instruction Manual

Table of Contents

Advertisement

4.2 Periodic Maintenance

No.
1
2
3
4
5
6

4.2.1 Liquid Leak Check

4.2.2 Washing the Flow Path

Part No.
S264000
775-1365
Estimated periodic maintenance is based on full operation of 8
hours daily and 20 days monthly with standard assay.
Check Item
Liquid leak check
Washing the flow path

Cleaning

Waste solution tube

Liquid leak sensor

Drain tray
Visually check that liquid is not leaking through any tube.
The flow path should be washed on a periodic basis. Bubbles
will be formed if the flow path between the syringe and needle is
contaminated; these bubbles will adversely affect the analytical
reproducibility of the system.
(1) When using an aqueous detergent
At the end of measurement, wash the flow path with distilled
water. If contamination is heavy, fill the syringe with
detergent and leave it overnight. After cleaning the syringe,
rinse it with distilled water. If the instrument will not be
used for a week or loner, wash the flow path adequately
with distilled water and keep the syringe filled with distilled
water. In particular, the syringe and sample loop should be
washed carefully to remove contaminants. And prevent the
inside of flow path from drying out. If the flow path dries
out though contaminated, decontamination may become
impossible.
Table 4-1
Detergent
(brand)
Extran®
20 to 50 times
Rinsing liquid
Not diluted
4 - 8
Frequency
Daily
As required
As required
Daily
Monthly
As required
Recommended Neutral Detergents
Concentratio
n
Dilute the detergent with ion-
exchanged water or distilled
water when using phosphorus-
free neutral detergent (Extran
MA03, etc.).
Procedure
Refer to 4.2.1
Refer to 4.2.2
Refer to 4.2.3
Refer to 4.2.4
Refer to 4.2.5
Refer to 4.2.6
Precaution

Advertisement

Table of Contents
loading

Table of Contents