Hitachi CHROMASTER 5160 Instruction Manual page 473

Table of Contents

Advertisement

(d) The following display will be presented.
<Em bandwidth selecting screen>
Em BANDWIDTH (STANDARD=1, WIDE=2)
1
Set an Em bandwidth value.
Press numeric key (1 or 2, usually input 1) and then
the ENT key.
* Default value: 1
(e) Em bandwidth setting is now completed and the 1
monitor screen returns.
(2)
ERROR OUT setting
It is selectable whether to enable or disable the error signal
output.
When enabled, the following errors can be indicated.
Xenon lamp error
Thermo cell error
Thermo cell connection error
ADC error
DAC error
Data buffer overflow (CDS system)
Overheat (Refer to Section 4.7)
(a) Press the
(b) The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_5
USE PROG=3, PMT VOLT=4, OTHERS=5)
Select OTHERS.
Press numeric key 5 and then the ENT key.
3 - 13
SET UP
UTLTY
key.
st

Advertisement

Table of Contents
loading