Hitachi CHROMASTER 5160 Instruction Manual page 882

Table of Contents

Advertisement

(continued)
2 ENT
SYRINGE CHANGE
<1-12>
SYRINGE CHANGE
<1-31>
SYRINGE CHANGE
<2000-2100>
Monitor screen reappears.
4 ENT
LOGBOOK (REPORT=1, RESET=2)
1
Select 1 for logbook display or 2 for resetting the
logbook.
1 ENT
S.VALVE SEAL
Monitor screen reappears.
(to be continued)
3 - 38
(MM DD YYYY)
1
Input the month of replacement.
6 ENT
(MM DD YYYY)
6
Input the date of replacement.
2 ENT
(MM DD YYYY)
Input the year of replacement.
Upon completion of all the inputs up to year,
the logbook is reset.
2010 ENT
LAST CHANGED DATE
23 times
This screen displays the number of syringe
valve switchings to port 1 after the latest
replacement of the syringe valve seal and the
date of the replacement.
ENT or ESC
1
2000
1 2000
6
2
2000
1/1/2000