Hitachi CHROMASTER 5160 Instruction Manual page 866

Table of Contents

Advertisement

(continued)
2 ENT
NEEDLE DOWN SPEED (SLOW=1, FAST=2)
2
Monitor screen reappears.
3 ENT
SYRINGE VOL (70 µL=1, 175 µL=2, 700 µL=3,
2
Monitor screen reappears.
4 ENT
AIR VOL (0.0-20.0) µL
2.0
Monitor screen reappears.
(to be continued)
3 - 22
Input a needle lowering speed. Usually, set
"FAST." For using a hard septum, "SLOW"
should be set.
Needle Down
Speed
SLOW
FAST
NOTE: When the slow needle down speed is
selected, the needle is apt to be
broken due to a strong falling force.
Therefore, before operation, make
sure vials, etc. are correctly set.
2 ENT
Select a syringe volume to be used.
2 ENT
Input an air volume.
2 ENT
Speed
30 mm/s
150 mm/s