(c) The following display will be presented.
<Others setting screen>
OTHERS (Em BANDWIDTH=1, ERROR OUT=2,
3
LEAK SENSOR=3, LAMP=4)
Select LEAK SENSOR.
Press numeric key 3 and then the ENT key.
(d) The following display will be presented.
<Leak sensor selecting screen>
LEAK SENSOR (OFF=0, ON=1)
1
Select use (ON=1) or non-use (OFF=0) of the leak
sensor.
Press a numeric key (0 or 1, usually input 1) and then
the ENT key.
* Default value: 0
* When use (ON=1) of the leak sensor is selected, an
error message will be indicated upon detection of
liquid leakage (Refer to Section 4.6).
* When power supply is restored, setting will be use
(ON=1) of the leak sensor.
(e) Selection or avoidance of the leak sensor is now
completed and the 1
(4)
LAMP setting
ON or OFF status of the xenon lamp is selectable in the
following procedure.
(a) Press the
(b) The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_5
USE PROG=3, PMT VOLT=4, OTHERS=5)
3 - 15
st
monitor screen returns.
SET UP
UTLTY
key.