2.4.2 Thermostat (option)

(1)
Temperature setup range*
1 to 45 °C (in increments of 1 °C)
(2)
Temperature control range*
[ambient temperature – 21 °C] to [ambient temperature + 25 °C]
and within the setup range
(humidity 60 %, sample vial 1.5 mL, 4 mL, 1 mL)
[ambient temperature – 15 °C] to [ambient temperature + 20 °C] and
within the setup range (humidity 60 %, microplate)
(3)
Temperature setting accuracy
±3 °C
(Target temperature 4 °C/40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central
sample vial (1.5 mL, 4 mL, 1 mL), without sample freezing)
±4 °C
(Target temperature 10 °C/40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central
microplate without sample freezing)
(4)
Temperature distribution
±2 °C
(Target temperature 4 °C /40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central
and 4-corner sample vials (1.5 mL, 4 mL, 1 mL), without
sample freezing)
±3 °C
(Target temperature 10 °C /40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central
and 4-corner sample vials (each microplate), without sample
freezing)
(5)
Maximum number of sample vials
120 (1.5 mL) (standard)
72 (4 mL) (option)
195 (1 mL) (option)
192 (96-holes microplate × 2) (option)
768 (384-holes microplate × 2) (option)
2 - 47