(4)
<Standard Xe lamp lifetime setting screen>
XE LAMP EXCHANGE STANDARD (1 - 2500) H
_500
Part No.
Part Name
650-1500
150 W Xe lamp
890-2019
Long-life Xe lamp
(5)
<Setup initial screen>
SETUP: DETECT
6)
Select EXCHANGE STANDARD.
Press numeric key 4 and then ENT key.
The following display will be presented.
The standard lifetime of the Xe lamp is settable here.
Press numeric key (to input 1 to 2 5 0 0) and then the ENT
key.
* Check the part number of your Xe lamp and input the
correct value listed below.
Guaranteed Life
150 hours
500 hours
* Default value: 500 hours
(J851152 Xe lamp is attached with shipment.)
The setting of a standard Xe lamp lifetime is now completed
and the following display will be presented.
THERMO
CH1
* When you press the ENT key here, the setup items
selecting screen is presented to allow reentry.
* The selected setting will be fetched upon turning on
power supply at the next time.
Turn OFF the power switch.
3 - 51
Limit
500 hours
1,000 hours
CONTRAST
USE
3
Allowable Input
500 or less
1000 or less
S. NO.
02AB-111