3.2 Utility Setting

3.2 Utility Setting

3.2.1 Setting the Offset

The utility setup screen appears.
UTILITY (OFFSET=1, TIME CONSTANT=2,
1
USE PROG=3, POLARITY=4, OTHERS=5)
Input Numeral
1
OFFSET
2
TIME CONSTANT :
3
USE PROG
4
POLARITY
5
OTHERS
The offset is a set input signal value that is added to the actually
measured value for output.
SET UP
UTLTY
1)
Press
2)
The utility setup screen appears.
Utility Setup Screen
UTILITY (OFFSET=1, TIME CONSTANT=2,
1
USE PROG=3, POLARITY=4, OTHERS=5)
Select OFFSET.
Press [numerical] key [1], and then the [ENT] key.
3)
The offset setup screen appears.
Offset Setup Screen
OFFSET (0.000 – 2.000) AU
0.000
3 - 10
SET UP
UTLTY (Utility)
Function
:
Set an offset.
Set a time constant.
:
Select or avoid use of a time
program.
:
Select either polarity.
:
Other setting than above.
key.