3.4 Analyzing Operation
During spectrum acquisition, the following screen is
presented.
SCANNING SPECTRUM NO.: 1 MODE: W REF
WL: *** nm
After spectrum acquisition, the monitor screen automatically
appears.
(e)
Storing the reference energy with mercury lamp ("5" key
followed by ENT key)
Upon selection of this function, the mercury lamp is ignited
and automatically selected during spectrum scan.
Set the minimum wavelength of scan range.
SCANNING WL RANGE (WL1 WL2) nm
<200 – 850>
Press [numerical] keys (within [2][0][0] to [8][5][0]), and then
the [ENT] key.
* Enterable range: 200 to 850 nm, enterable in 1-nm
steps
Set the maximum wavelength of scan range.
SCANNING WL RANGE (WL1 WL2) nm
<250 – 600>
Press [numerical] keys (within [2][5][0] to [9][0][0]), and then
the [ENT] key.
* Enterable range: Minimum scan range is 50 nm.
Measuring span is max. 400 nm.
Input the number of spectral data.
MEMORIZE SPECTRUM NO.(1–3)
_
3 - 34
Hg
200 – 600
Hg
200 – 600