Hitachi CHROMASTER 5160 Instruction Manual page 132

Table of Contents

Advertisement

4.4.2 Abnormal Pressure
Symptom
Abnormal high pressure
Abnormal variation of pressure
Check position
Possible cause
Clogged Column Filter.
Column
or Clogging caused by
contamination of Fillers.
Clogging of
Drain valve filter.
Flow path
system
Clogging of piping
between Drain
valve and Column
lnlet.
Bubbles are
generated in
Check valve.
Check valve
Foreign matters in
Check valve.
Life of Pump seal.
Pump seal
(Leakage)
Clogging of Drain
Flow path system
valve filter.
Bubbles in Pump
inlet tube.
Solvent filter
Contamination of
Solvent filter. (Clogging)
Check
Remove the Column, and pump the liquid.
Has the pressure become close to zero?
Remove the filter, and clean it with an ultrasonic cleaner.
Has the pressure become high?
Remove the piping one by one from the Column side, and find
an abnormal point.
Result
Remedy
Yes
Replace the column.
No
OK
Yes
Replace the filter. (See 5.5.)
No
OK
(1) Attach the piping in the reverse direction, perform pumping with
a high flow rate to remove foreign matters.
(2) Cut the piping from the inside of the Ferrule part, and use it
again.
(3) After using a salt solvent, substitute distilled water for the inside
of the Flow path system.
Open the Drain valve fully, pump liquid for 3-5 minutes with
the maximum flow rate. (This should be performed when
bubbles are generated in the Pump after leaving the Pump
for a long time (2-3 days or longer).)
Open the Drain valve fully, pump liquid for 3-5 minutes with
the maximum flow rate.
Replace the seal. (See 5.3.)
See "Abnormal high pressure  Flow path system".
Degassing solvent
Cleaning with an ultrasonic cleaning, or replace the filter.
4 - 16