Use Time Program Setting - Hitachi CHROMASTER 5160 Instruction Manual

Table of Contents

Advertisement

3.2.3 Use time program setting

Select TIME CONSTANT.
Press numeric key 2 and then the ENT key.
(3)
The following display will be presented.
<Time constant setting screen>
TIME CONSTANT (0.01s=1, 0.02s=2, 0.05s=3,
4
Set a time constant value.
Press numeric key (1 to 7, usually input 4) and then the
ENT key.
* Default value: 4
(4)
Constant setting is now completed and the 1
screen returns.
It is selectable whether or not to use automatic change of
measurement parameters in response to measurement time
(time program).
SET UP
UTLTY
(1)
Press the
(2)
The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_3
USE PROG=3, PMT VOLT=4, OTHERS=5)
Select USE PROG.
Press numeric key 3 and then the ENT key.
(3)
The following display will be presented.
<Time program setting screen>
USE TIME PROGRAM (YES=1, NO=0)
1
3 - 10
0.1s=4, 0.5s=5, 1.0s=6, 2.0s=7)
key.