3.2 Installation Space
4. Check against Packing List
For composing a liquid chromatograph system with the
Chromaster series module units, a space of 540 mm wide
(980 mm for stacking the units in two blocks) and 550 mm deep
is required at least. Behind each unit, a space of 150 mm or
more must be secured. At the left and right sides of the
respective units, a space of minimum 100 mm each is necessary.
For composing the system in two-block stacking, the module
units should be positioned so that the clearance between the two
blocks becomes 100 mm. If the clearance between the blocks
is wider than 100 mm, the length of DC power cable may
become inadequate.
For installation, prepare a level and sturdy bench which
withstands a load of 150 kg or more. The system conceived
here is an example in which Chromaster organizer, Chromaster
5110 pump, Chromaster 5210 autosampler, Chromaster 5310
column oven and Chromaster 5410/5420/5430/5450 detector are
stacked. Confirm the necessary space taking the actual system
configuration and OA equipment in the installation room into
account and secure an adequate space at the time of installation.
Check the delivered items against the packing list.