Explained here is how to create a new time program
exemplifying the one shown above.
1)
At start, the measuring wavelength is set at 250 nm
and auto zero adjustment is carried out for baseline.
2)
After 10 minutes, the measuring wavelength is
changed to 270 nm. The baseline holds the previous
status (hold status).
3)
After 20 minutes, the time program ends.
PROG
Press
key.
SET PROGRAM NO.
1
Press [numerical] key [1], and then the [ENT] key.
EDIT MODE (MODIFY=0, NEW=1)
0
Press [numerical] key [1], and then the [ENT] key.
TIME
WL
0.0
250
Press [numerical] key [0], and then the [ENT] key.
Press [numerical] keys [2][5][0], and then the [ENT] key.
Press [numerical] key [1].
TIME
WL
0.0
250
Press the [ENT] key.
TIME
WL
_
Press [numerical] keys [1][0], and then the [ENT] key.
Press [numerical] keys [2][7][0], and then the [ENT] key.
Press numerical] key [2].
(Continued)
3 - 41
(1–9)
BASE(A/Z=1, HOLD=2)
1
BASE (A/Z=1, HOLD=2)
1