3.1.7 Programming Operation

PROGRAM N0. = 1
PRESSURE VALUE :
TIME
(min)
0.0
10.0
Operation with a program for isocratic mode is explained here.
As an example: upper pressure limit at 20.0 MPa, lower
pressure limit at 0 MPa, and flow rate is1.0 ml/min at 0 minute of
analysis time. Output pulse signals by Event 1 to 4.
Change the flow rate to 0.5 ml/min by 10 minutes of analysis
time.
The parameters in the table below are an example of input.
MAX
MIN
20.0
0.0
FLOW
(ml/min)
1
1.000
12
0.500
3 - 13
MPa
EVENT
2
3
4
22
32
42