Hitachi CHROMASTER 5160 Instruction Manual page 378

Table of Contents

Advertisement

(5)
Check the energy of W lamp. (Refer to 4.4.1.)
(6)
Input the date of W lamp replacement as follows.
(a) Press the (GLP) key, and the following screen will
appear.
GLP (LOGBOOK=1, KEY LOCK=2,
1
D2 CHECK=3, W CHECK=4, Hg CHECK=5)
(b) Press [numerical] key [1], and then the [ENT] key.
The following screen appears.
LAMP SELECT (D2=1, W=2, Hg=3)
1
(c) Press [numerical] key [2], and then the [ENT] key.
The following screen appears.
LOGBOOK (REPORT=1, RESET=2)
1
(d) Press [numerical] key [2], and then the [ENT] key.
The following screen appears.
W LAMP CHANGE (MM DD YYYY)
<1–12>
_9
(e) Key in the date of replacement in the format of month-
day-year. For instance, if the lamp is replaced on
March 1, 2010, key in the date as shown below.
[3] → [ENT] → [1] → [ENT] → [2][0][1][0] → [ENT]
Thus, the date of lamp replacement is stored in
memory. At the same time, the cumulative turn-on
time of the W lamp is cleared to zero (0), and the turn-
on count is reset to 1.
NOTE: The logbook data is reset when the numeric value
of the year and [ENT] are keyed in.
Therefore, even if the same year as that of
replacement is indicated, the year must be
numerically input followed by pressing the [ENT]
key.
5 - 11
15
2010
5.2.2