View from top
(Instrument front)
Hg light
entrance
Flow cell
(3)
Assembly of Flow Cell
(a) Assemble the flow cell into the cell holder in the
orientation shown in Fig. 4-11 and fix it with the cell
clamp.
(b) Tighten the upper and lower retaining screws 1 and 2.
Reflection
plate
(Instrument rear)
Fig. 4-13 Assembly of Flow Cell Unit
(c) Attach the cell window and shielding plate.
(d) After assembling the flow cell in the order reverse to
disassembly, flow a liquid through the unit and make
sure there is no leakage.
NOTE: Be careful not to scratch the flow cell surface when
NOTE: Take care about the orientation of the polished
NOTE: Tighten setscrew 1 fully with hand, and then
Enlarged view of flow cell
(Instrument front)
Transmitted
excitation light
Emission
(Instrument rear)
assembling it.
surfaces of the flow cell when putting it into the cell
holder (see Fig. 4-11). A mistake in the
orientation may preclude obtaining the proper
performance from the instrument.
tighten setscrew 2 about 30° manually or tighten
setscrew 1 further 30° with a wrench.
4 - 28
Polished surface
(without slit)
Excitation light
Polished surfaces
(with slit)