3.2 UTLTY Setting
3.2 UTLTY Setting
3.2.1 Setting of Leak Sensor
SET UP
UTLTY
UTILITY (LEAK SENSOR=1, ANALOG OUT=2,
1
ERROR OUT=3, CALIBRATION=4)
SELECT SENSOR (GAS=1, LIQUID=2)
1
LEAK GAS (MONITOR=1, SET LEVEL=2)
1
SENSOR MONITOR
400
Monitor screen
(1)
Set the leak detection level for the gas leak sensor.
1 ENT
1 ENT
Monitor the gas leak sensor
output.
1 ENT
ALARM LEVEL
1200
ENT or ESC key
SET ALARM LEVEL (0 - 4095)
Set up a gas leak detection level.
1200 ENT (initial value: 1200)
3 - 10
1200
Monitor screen
2 ENT