Shutdown Procedure; Flushing The Flow Cell After Use - Hitachi CHROMASTER 5160 Instruction Manual

Table of Contents

Advertisement

3.1.5 Shutdown procedure

3.1.6 Flushing the flow cell after use

Then, the cursor moves to Em.
(3)
Set an emission wavelength (Em).
<Em wavelength setting screen>
Ex
Em
250
350
Press numeric keys ((2 5 0 to 9 0 0)) and then the ENT key.
* This wavelength is settable within a range from 250 to
900 nm in increments of 1 nm.
* Default value: 350 nm
* Press the ESC key or
emission wavelength value. It returns to the 1
screen.
* Pressing the
excitation wavelength setting without changing the
emission wavelength.
(4)
Wavelength setting is now completed and the 1
screen returns.
At Ex and Em, the set values are indicated.
st
<1
monitor screen>
TIME
Ex
Em
0.0
250
350
Turn the power supply of the detector off to shut it down.
The set measurement conditions are retained by the memory
even when the power is switched off. When power is turned on
again, the conditions that were in place when power was turned
off will be restored.
Make sure to flush the flow cell with distilled water after using
any buffers or salts in the mobile phase. Failure to carry out
flushing may lead to clogging and/or breakage of the flow cell.
3 - 7
(250 - 900.0) nm
key to cancel the input
key enables you to return to the
FL
PROG
0.0
1
st
monitor
st
monitor
LAMP
[I]
Xe