4.1.3 Reproducibility Check

(13) Input the following contents as instrument conditions.
RACK CODE:1
SYRINGE SPEED:1
RACK PARAM:RACK CODE = 1
X1 = 57.1
X2 = 178.6
Nx = 10
Y1 = 0.0
Y2 = 148.5
Ny = 12
Z = 34
P = 5
INJ.METHOD: CUT
LEAD VOLUME: 30.0
REAR VOLUME: 30.0
ASP SYRINGE SPEED: 3
DSP SYRINGE SPEED: 2
AIR VOLUME: 2.0
NEEDLE WASH TIME SOLVENT1: 4
RINSE PORT WASH TIME: 1
SYRINGE VOL:2 (175 µL)
(14) Set the pump flow rate to 1.0 mL/min.
(15) Set the detector wavelength to 265 nm. Set the detector
time constant to 0.1 s.
(16) Set the following conditions in the data processing unit.
SAMPLE PERIOD : 100 ms
STOP TIME
Check the reproducibility of the injection volume in the following
procedure.
(1)
Press the PROG key and set the following conditions.
STEP RACK
1
1
Pressing the ESC key can complete setting.
(2)
Prepare 60 % methanol for the mobile phase and wash
fluid. Wait until the pump, detector, etc. are warmed up
and the baseline stabilizes.
4 - 5
: 3 min
VIALS
VOL(µL)
1- 1
10.0
INJ
TIME
6
3.0