(c) The following display will be presented.
<Others setting screen>
OTHERS (Em BANDWIDTH=1, ERROR OUT=2,
2
LEAK SENSOR=3, LAMP=4)
Select ERROR OUT.
Press numeric key 2 and then the ENT key.
(d) The following display will be presented.
<Error out signal selecting screen>
ERROR OUT SIGNAL (DISABLE=0, ENABLE=1)
1
Invalid (0) or valid (1) status of error out signal output is
settable.
Press numeric key (0 or 1) and then the ENT key.
* Default value: 1
(e) Error out setting is now completed and the 1
screen returns.
(3)
LEAK SENSOR setting
Whether to use the liquid leak sensor or not is settable.
(a) Press the
(b) The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_5
USE PROG=3, PMT VOLT=4, OTHERS=5)
Select OTHERS.
Press numeric key 5 and then the ENT key.
3 - 14
SET UP
UTLTY
key.
st
monitor