Hitachi CHROMASTER 5160 Instruction Manual page 114

Table of Contents

Advertisement

Devices to be prepared
(i)
Ultrasonic cleaner
(Put water in the bath without fail.)
(ii) Aspirator with rubber stopper which fits the mouth of
mobile phase container
Rubber stopper
should fit the mouth
of container.
Mobile phase container
Fig. 3-9 Degassing Mobile Phase by the Aspirator
(a) Set mobile phase container in ultrasonic cleaner after
sufficient stirring.
* When using a mixed mobile phase, stir it
completely by using a stirrer, or else poor
repeatability or drift of baseline may be caused.
NOTE: Degas mobile phase of aqueous solution by using both
ultrasonic cleaner with water and aspirator to decrease
pressure. Fix the degassing time.
(b) Insert rubber tube of aspirator into mouth of reagent
bottle.
(c) Turn on the power of ultrasonic cleaner.
(d) Open the faucet of city water for aspirator to decrease
pressure.
(e) When degassing, small bubbles are generated first,
then gradually larger bubbles are generated.
Continue degassing until bubble generation finishes.
It takes about 5 minutes. (See NOTE.)
3 - 77
Ultrasonic cleaner
Aspirator

Advertisement

Table of Contents
loading

Table of Contents