(b) 1st Monitor Screen (Oven Temperature Indication)
TEMP:MONITOR SET
25.1
Indication
MONITOR
Measured temperature of column oven
SET
Set temperature of column oven
RANGE
Range in which temperature becomes ready
(set temperature ± tolerance)
WAIT
Waiting time until temperature ready is attained after
entering RANGE ("–" indicated if waiting time is '0')
Description of status indication at upper right of monitor
screen:
[I]............. Oven temperature is in ready status.
[R] ........... Run status (Time program is running)
[B] ........... Oven temperature is in busy status.
[L]............ Keyboard is locked.
[–]............ Status other than above (temp. control stopped)
(c) 2nd Monitor Screen (Ambient Temperature Indication)
TEMP:AMBIENT
25.0
Indication
AMBIENT
Indication of ambient (room) temperature
VALVE1
Indication of present position of selector valve (option)
"–" indicated if valve is not connected.
Refer to (6) for the description of status indication at upper
right of monitor screen.
3 - 3
READY:RANGE
40
39.5-40.5
Description
VALVE 1
1
Description
3.1.1
WAIT [-]
5.0
[-]