3.2.4 PMT voltage setting

Select use (YES=1) or non-use (NO=0) of a time program.
Press a numeric key (1 or 0) and then the ENT key.
* 1: Time program starts by start input.
0: Time program does not start despite start input.
* Default value: 1
(4)
Selection or avoidance of a time program is now completed
st
and the 1
monitor screen returns.
When use of a time program is avoided, "-" is indicated for
PROG.
<1st monitor screen>
TIME
Ex
Em
0.0
250
350
Set photomultiplier voltage for fluorescence detection.
SET UP
UTLTY
(1)
Press the
(2)
The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_4
USE PROG=3, PMT VOLT=4, OTHERS=5)
Select PMT VOLT.
Press numeric key 4 and then the ENT key.
(3)
The following display will be presented.
<PMT voltage setting screen>
PMT VOLT (SUPER HIGH=1, HIGH=2, MID=3,
3
LOW=4, SUPER LOW=5)
3 - 11
FL
PROG
0.0
key.
LAMP
[I]
Xe

Advertisement