Hitachi CHROMASTER 5160 Instruction Manual page 299

Table of Contents

Advertisement

(c)
Setting the Use (ON) or Non-use (OFF) Status of Leak
Sensor
SET UP
UTLTY
1)
Press
2)
The utility setup screen appears.
Utility Setup Screen
UTILITY (OFFSET=1, TIME CONSTANT=2,
1
USE PROG=3, POLARITY=4, OTHERS=5)
Select OTHERS.
Press [numerical] key [5], and then the [ENT] key.
3)
The OTHERS setup screen appears.
OTHERS (PROCESSOR=1, ERROR OUT=2
1
LEAK SENSOR=3, LAMP=4, GAIN=5)
Select LEAK SENSOR.
Press [numerical] key [3], and then the [ENT] key.
4)
The LEAK SENSOR setup screen appears.
LEAK SENSOR (OFF=0, ON=1)
1
Set the use (ON) or non-use (OFF) status of LEAK
SENSOR.
Press [numerical] key ([1] or [0]), and then the [ENT]
key.
* Default value: 1
5)
Upon completion of LEAK SENSOR setting, the first
monitor screen reappears.
NOTE: If power supply is turned off with the leak sensor
set at OFF, you should remember that the "ON"
setting returns upon turning on power supply
again.