3.2.3 Setting the Use or Non-use of Time Program
Automatic change of measurement conditions with time (time
program) can be selected or avoided.
SET UP
UTLTY
1)
Press
key.
2)
The utility setup screen appears.
Utility Setup Screen
UTILITY (OFFSET=1, TIME CONSTANT=2,
1
USE PROG=3, POLARITY=4, OTHERS=5)
Select USE PROG.
Press [numerical] key [3], and then the [ENT] key.
3)
The time program setup screen appears.
Time Program Setup Screen
USE TIME PROGRAM (YES = 1, NO = 0)
1
Set use (1) or non-use (0) of the time program.
Press [numerical] key ([1] or [0]), and then the [ENT] key.
* 1 : Time program starts in response to start input.
0 : Time program does not start irrespective of start
input.
* Default value: 1
4)
Upon completion of time program selection or avoidance,
the first monitor screen reappears.
When "0" is set, "-" appears at PROG.
First Monitor Screen
TIME
WL
0.0
250
3 - 13
ABS
PROG
LAMP
0.0000
–
D2
3.2.3