3.2 UTLTY Setting

3.2.1 Zero Pressure Calibration

The zero pressure point may change a little due to a secular
change of the pressure sensor. Please check the zero
pressure point once a month (Stop the pump, and then open the
drain valve). Refer to the following operation for calibration.
Monitor screen
SET UP
UTLTY
UTLITY
(CONDITION=1, UNIT=2,
1
EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
Select CONDITION
1
ENT
CONDITION SET
1
PURGE=3,
Select Z-ADJ (ZERO ADJUST)
1
ENT
PRESSURE ZERO ADJUT
1
(NOTE: OPEN DRAIN VLAVE)
Select YES
1
ENT
Monitor screen
PRESS is displayed as 0, and the zero pressure calibration is
completed.