(6)
Setup Processing
You can select use or nonuse of the thermostat (option),
and set LCD contrast of the UI pad.
(a) Selection of use or nonuse of thermostat (option)
+
POWER ON
Turn on power supply with the UTLTY key held down.
SETUP: THERMO UNIT CONTRAST S.NO.(A/S)
NO USE
ENT
SETUP ITEMS (THERMO UNIT=1, CONTRAST =2)
1
1 ENT
THERMO UNIT (NO USE=0, USE=1)
1
Enter 0 for avoiding use of the thermostat, and 1 for
selecting its use.
1 ENT
SETUP: THERMO UNIT CONTRAST S.NO.(A/S)
USE
ENT
Turn off power supply
and turn it on again.
When turning on power supply again, the instrument starts
under the set conditions.
3 - 51
UTLTY
(SET UP)
4
02AB-111
Indicates a serial No. (product No.).
4
02AB-111